|
|
||||||||
Imaging & Therapeutic Technology |
1 From The Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, 601 N Caroline St, Baltimore, MD 21287. From the AAPM/RSNA Physics Tutorial at the 1999 RSNA scientific assembly. Received February 16, 2001; revision requested March 29 and received April 19; accepted April 23. Address correspondence to the author (e-mail: mmahesh@jhmi.edu).
| Abstract |
|---|
|
|
|---|
Index Terms: Dosimetry Education Fluoroscopy Physics Radiations, exposure to patients and personnel Radiations, injurious effects Radiations, measurement
| Introduction |
|---|
|
|
|---|
In response to the problem, the Center for Devices and Radiological Health of the Food and Drug Administration (FDA) issued an advisory in 1994 (2) warning health care facilities of the potential for radiation-induced burns to patients from prolonged fluoroscopic procedures. According to the advisory, a number of interventional procedures (Table 1) have the potential to cause skin injury even when the fluoroscopic time is 1 hour or less at normal dose rates. An insidious aspect of the problem is that onset of injury is delayed and the extent of damage may not be evident until weeks after the procedure. Approximately 50 injuries were reported to the FDA in 1994 alone (1), and since then the FDA has documented many more cases of radiation-induced burns (3). A number of case histories of injuries to both patients (47) and physicians (8) have subsequently appeared in the literature; some of the radiation-induced wounds have required skin grafts, resulting in permanent disfigurement. The actual extent of the problem is essentially unknown, since there are currently neither requirements for reporting nor a central repository for this information.
|
| Biologic Effects of Radiation |
|---|
|
|
|---|
On the other hand, there are other effects for which the probability of causing certain types of harm will be zero at small radiation doses. Above some threshold level, damage will become apparent, with severity increasing as dose rises above the threshold. Such effects are called nonstochastic or deterministic effects. Cataracts, erythema, epilation, and even death are examples of the deterministic effects that can result from high radiation exposures.
Except in the early days of radiology, when precautions were few, the emphasis in radiation safety has been on stochastic effects. The recent popularity of prolonged interventional procedures has added concern about deterministic effects. Since, at diagnostic x-ray energies (eg, those for fluoroscopy, radiography, computed tomography, and angiography), doses are highest at the beam entrance point, the most commonly observed harm has been skin tissue damage and hair loss. Table 2 summarizes some of the skin reactions and their dose thresholds (9).
|
Although other factors are important, the performance of a fluoroscopy system with respect to radiation dose is best characterized by the receptor entrance exposure rates and skin entrance exposure rates. In addition, exposure rates are dependent on the patient thickness and operational factors; hence, it is becoming increasingly apparent that some sort of dynamic patient dose monitoring is desirable during extended fluoroscopic procedures.
| Receptor Entrance Exposure Rates |
|---|
|
|
|---|
Receptor entrance exposure is normally specified as the entrance exposure at the surface of the image receptor (with the grid removed) required to produce a single image for a given x-ray spectrum. In the measurement geometry, the ionization chamber is placed 2030 cm from the image intensifier surface (with a fixed source-toimage intensifier distance of 100 cm) to reduce the backscatter contribution (11) (Fig 1). The exposure rates are corrected to the entrance surface of the image intensifier by using the inverse square law. Receptor entrance exposure rates are measured with the grid removed (12); however, it is often impractical to remove the grid from some systems, in which case the manufacturer-specified grid transmission factor can be used to correct for the presence of the grid (11).
|
|
|
| Skin Entrance Exposure Rates |
|---|
|
|
|---|
The maximum skin entrance exposure value is important, but to obtain a realistic estimate of the level that a patient would receive, a measurement of the average skin entrance exposure is required. Skin entrance exposure values can vary widely with x-ray tube voltage (kilovolt peak) and patient thickness, presence or absence of a grid, source-to-image distance, and so on. As shown in Table 5, an hour of exposure with typical skin entrance exposure rates during normal fluoroscopy and high-dose fluoroscopy can yield a patient entrance exposure of 0.63 Gy and 612 Gy, respectively. When these values are compared with threshold levels associated with skin injuries (Table 2), it is clear that the potential exists for skin injuries during prolonged fluoroscopic procedures.
|
| Patient Dose Monitoring Methods |
|---|
|
|
|---|
There are currently a number of methods available by which one can estimate or measure patient skin dose; they may be classified as direct or indirect methods. Direct methods involve measurement of skin dose during the procedure by using one of several types of small dosimeters taped to the patients skin. Indirect methods involve calculation of skin dose from measurements in the beam or from operational factors that affect dose.
Direct Methods
The direct method of skin dose estimation involves use of small detectors placed on the patients skin at the beam entrance location. Several types of detectors have been used, including thermoluminescent dosimeters (TLDs), photographic films (1619), and, more recently, diodes or metal-oxide semiconductor field-effect transistor (MOSFET) detectors. Use of a TLD is potentially the most accurate way of determining actual skin dose, but results can be very inaccurate because it is often impossible to know exactly where on the skin the peak dose will occur in a given procedure. Use of an array of dosimeters can reduce uncertainty, but this strategy is not always practical. With single dosimeters, considerable care is required to correctly place the dosimeter and to ensure that it is not moved from the field during the procedure. The method requires that the entrance field be known a priori, and, since the x-ray source is usually located beneath the patient table, it requires the patient to lie directly on the detector during the procedure.
Photographic films have several advantages such as low cost, an easy-to-locate high-dose region, and dose measurement by using densitometers. However, there are only a few films available in large sizes that have the sensitivity to measure several ranges of doses. MOSFET detectors have the advantage of providing a dynamic reading of the skin dose as it accumulates during the procedure, but some physicians find them to be objectionable due to the visibility of the detector elements and connecting leads in the image field. Although TLDs are essentially invisible, they are read out after the procedure and cannot provide a dynamic display of dose during the procedure.
Indirect Methods
The most convenient and widely used method for indirect monitoring is the dose-area product (DAP) meter. The DAP meter uses a transmission type air-ionization chamber mounted on the face of the x-ray tube collimator, which integrates exposure over the entire image field. The DAP measurement is a function of the x-ray field size and the x-ray exposure at the collimator; thus, the measurement is expressed as either the dose-area product or the air-kerma-area product (2024). The measured DAP is independent of distance from the focal spot. The distance factor cancels because the exposure rate varies inversely and the x-ray field area varies inversely as the square of the distance from the focal spot to the point of measurement. A given DAP reading can result from a high dose over a small field or a low dose over a large field. The stochastic risk can be roughly assumed to be equivalent under these two conditions; thus, DAP meters have been used to estimate total stochastic risk for latent effects (14,15). Unfortunately, a high dose over a small field is not equivalent to a low dose over a large field in terms of skin damage; hence, the DAP reading requires some interpretation.
Skin entrance dose can be computed from DAP readings only at a specific dose rate and field size. For example, if a 1-minute fluoroscopic procedure produces a DAP reading of 2,000 µGy · cm2, this result could be obtained with a 20 x 20-cm field at 5 µGy/min or a 10 x 10-cm field at 20 µGy/min, a factor of four difference in skin dose. When magnification mode is selected, the input field size is reduced, which results in decreased brightness. The automatic brightness stabilizers on many systems then respond by increasing the x-ray dose to maintain constant image brightness. This process can produce a static DAP reading (Fig 2), even though the skin dose rate scales by a factor of two or more. In situations where dose rate and field size are changing dynamically, accurate estimation of skin dose from DAP readings may be problematic.
|
There is at least one system available that will dynamically record and display an estimate of total skin entrance exposure. This product is based on the fact that skin dose can be predicted from operational factors. This device, currently marketed under the name PEMNET (Clinical Microsystems, Arlington, Va), measures peak kilovoltage, tube current (milliamperage), and exposure time signals from the x-ray generator to dynamically compute and display an estimate of the skin entrance dose based on calibration of radiation output as a function of these parameters (25,26). Some versions have the capability to compensate for attenuation by the table and adjust for oblique angles (26) while providing skin entrance exposure at known distances. However, these systems share a limitation with DAP meters in that dose estimates do not take into account overlapping beams from several sources. The source-toskin entrance distance is a major uncertainty in the estimate, but some versions incorporate a proximity detector to measure this parameter during the procedure. The complexity of installation required for calibration has hindered widespread use of the system, although it has considerable potential.
Recent reviews by Geise and ODea (16) and Strauss (10) have examined the features of currently available skin dosimetry methods and have concluded that currently there is no perfect dose monitoring system for complex interventional procedures.
| Patient Doses for Typical Fluoroscopic Procedures |
|---|
|
|
|---|
3 to 15 minutes (18,20,21) with entrance skin dose ranging from 44 to 340 mGy (4.434 rad). Doses for the more complex interventional procedures are listed in Table 7. Diagnostic procedures performed under fluoroscopic control have significantly shorter fluoroscopic times and lower entrance skin doses compared with those of interventional procedures. Mean fluoroscopic times for interventional procedures are significantly longer (17,20,21).
|
|
|
2 Gy) was exceeded in 22% of procedures.
|
| Dose Reduction Techniques |
|---|
|
|
|---|
Intermittent Fluoroscopy
Most radiologists are trained to control the fluoroscope intermittently, that is, keeping the x rays on only a few seconds at a time, long enough to view the current catheter position. Judicious use of the method can reduce total fluoroscopic times considerably. This simple technique is particularly effective when combined with last image hold features.
Removal of Grid
The presence of grids in x-ray systems primarily increases the contrast and hence the image quality; however, they increase the dose to the patient and staff by a factor of two or more. Studies have shown that, especially in pediatric cases, removal of the grid has resulted in dose reduction of up to one-third to one-half with little or no degradations in contrast and image quality (30,31). Grids should be used with discretion when fluoroscopic examinations are performed on children, and the systems for such examinations should have the capability for easy removal and reintroduction of the grid.
Last Image Hold and Electronic Collimation
A useful feature on many modern fluoroscopy systems is last image hold, whereby the last image is digitally "frozen" on the monitor after x-ray exposure is terminated. Last image hold is a dose-saving feature (32), since it allows physicians to contemplate the last image and plan the next move without additional radiation exposure in an interventional procedure. In addition, some modern systems have electronic collimation, which overlays a collimator blade on the last image hold so that one can adjust field dimensions without exposing the patient.
Dose Spreading
In most interventional fluoroscopic procedures, the bulk of the fluoroscopic time is spent at a particular anatomic region during the procedure. For example, in radio-frequency ablation procedures, the fluoroscope is used to guide the catheter from the femoral artery to the heart but thereafter remains over the heart region. Some reduction of maximum skin dose can be achieved by periodically rotating the fluoroscope about a center within the anatomy of interest. This method tends to spread the maximum dose over a broader area of the patients skin so that no single region receives the entire dose (Fig 3).
|
Substantial reductions in skin dose are also achieved by inserting appropriate metal filters (aluminum, copper, or other materials) into the beam at the collimator. Filtration reduces skin dose by preferentially removing lower-energy photons, which would not penetrate the patient to contribute to the image. Nicholson et al (33) have shown that the addition of 0.10.3 mm of copper reduced the skin dose by 30%50%. In general, if optional filters have been designed into the fluoroscopy system (eg, Carefilters [Siemens Medical Systems, Iselin, NJ] and Spectrabeam [Philips Medical Systems, Shelton, Conn]), they should be used for all lengthy or repeated examinations provided the image quality is diagnostically acceptable.
Image Magnification
The ability to create magnified images can be clinically very useful but in almost all cases results in a higher patient dose. There are two basic ways to magnify the image in fluoroscopy: geometric and electronic. Geometric magnification takes advantage of the diverging x-ray beam to project a smaller region in the patient to a larger area on the image intensifier. When source-toimage receptor distance is fixed, both image magnification and skin dose increase as the patient is moved closer to the x-ray source (Fig 4). Many interventional system fluoroscopes do not fix the source-toimage receptor distance, and the positions of the tube and the receptor can both be changed independently. Thus, moving the source closer to the patient or the receptor further away can magnify the image. Also, there is increased penumbra (focal spot blur) with higher magnification, and unless a very small focal spot is used (eg, 0.3 mm), the spatial resolution is degraded. It is generally best to minimize geometric magnification in prolonged procedures by keeping the image receptor close to the patient and the source away.
|
|
Pulsed Fluoroscopy
Some modern fluoroscopes have the capability of pulsed fluoroscopy, whereby the x-ray beam is emitted as a series of short pulses rather than continuously. At reduced frame rates, pulsed fluoroscopy can provide substantial dose savings. Images may be acquired at 15 frames per second rather than the usual 30 frames per second. Each image is displayed multiple times in sequence to provide a 30 frames per second display. Pulsed fluoroscopy can also be performed at even lower frame rates (eg, 7.5 or 3 frames per second) at the expense of a "choppy" display when imaging rapidly moving regions like the heart.
Because simply reducing the number of pulses would result in an increase in image noise, manufacturers may increase the milliamperage setting to achieve a similar visual appearance. For example, one would expect a 50% dose reduction when going from 30 to 15 frames per second, but, because of increased milliamperage, the actual dose savings are 25%28% (Fig 6). With equivalent perceptibility levels, Aufrichtig et al (34) showed average dose savings of 22%, 38%, and 49% at 15, 10, and 7.5 frames per second, respectively. When operating at lower frame rates and higher magnification (decreased field of view), some systems open the television camera aperture instead of increasing the exposure to maintain similar image brightness. Pulsed fluoroscopy has a great advantage as long as the radiation exposure is lower at lower frame rates. If the tube current is set too high to achieve better-quality images, the entire advantage of pulsed operation is defeated and there may be no actual dose savings.
|
| Training of Fluoroscopic Operators |
|---|
|
|
|---|
To date, there are no uniform national standards regarding who may operate fluoroscopy systems or what minimum training they should have (3,36). All resident physicians undergo a credentialing process that involves verification and assessment of their qualifications to ensure that they are appropriately trained to perform clinical procedures with technical proficiency and henceforth grant clinical privileges (3,36). However, none of the credentialing processes mandate that resident physicians, especially those who work with radiation, learn about radiation management in fluoroscopic procedures (36). Except for radiologists and nuclear medicine specialists, most physicians receive minimal to no training in the safe use of ionizing radiation during their residency program. In addition, except for a few boards such as the American Board of Radiology, American Board of Nuclear Medicine, and Royal College of Physicians and Surgeons of Canada (diagnostic radiology), no other medical specialty board includes radiation in the curriculum (36). However, there have been positive developments, as seen in the position statements taken by some prominent organizations such as the American College of Cardiology (37) and American Heart Association (38), strongly advocating education of fluoroscopy users. For example, the American College of Cardiology consensus document, "Radiation Safety in the Practice of Cardiology" (37), clearly stresses the importance of radiation safety and education for everyone using fluoroscopy.
Some state regulations include mandatory education and training in the safe use of x rays. Even if this is not required by law, it is to the advantage of a facility to establish minimum requirements for the safe use of fluoroscopic equipment for all physicians performing fluoroscopy, particularly those performing prolonged interventional procedures. Any training or credentialing process should include education on radiation safety issues, proper use of fluoroscopic equipment, and control settings (35). In addition, biologic effects of radiation with emphasis on deterministic effects should be discussed along with detailed discussions on dose-reducing techniques. The training or credentialing process can be achieved by means of World Wide Webbased short courses, workshops, or tutorials with tests (39,40); by adding study materials to respective board curricula; or by developing institution-based courses. There are a number of sources, such as medical physics textbooks, AAPM reports (35), and medical literature in radiographic physics series, that can help a facility establish a program to train or credential their physicians using fluoroscopy to obtain minimal competency in terms of the safe use of radiation. Enlisting a qualified medical physicist to assist in training fluoroscopy users, as recommended in the FDA advisory (2), can be of great advantage to the facility.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. P. Kim and D. L. Miller MINIMISING RADIATION EXPOSURE TO PHYSICIANS PERFORMING FLUOROSCOPICALLY GUIDED CARDIAC CATHETERISATION PROCEDURES: A REVIEW Radiat Prot Dosimetry, March 27, 2009; (2009) ncp052v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A CHAFFINS Radiation Protection and Procedures in the OR Radiol. Technol., May 1, 2008; 79(5): 415 - 428. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Sulieman, K Theodorou, M Vlychou, T Topaltzikis, D Kanavou, I Fezoulidis, and C Kappas Radiation dose measurement and risk estimation for paediatric patients undergoing micturating cystourethrography Br. J. Radiol., September 1, 2007; 80(957): 731 - 737. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ector, O. Dragusin, B. Adriaenssens, W. Huybrechts, R. Willems, H. Ector, and H. Heidbuchel Obesity Is a Major Determinant of Radiation Dose in Patients Undergoing Pulmonary Vein Isolation for Atrial Fibrillation J. Am. Coll. Cardiol., July 17, 2007; 50(3): 234 - 242. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al. HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society. Europace, June 1, 2007; 9(6): 335 - 379. [Full Text] [PDF] |
||||
![]() |
L. D'Ercole, L. Mantovani, F. Z. Thyrion, M. Bocchiola, A. Azzaretti, F. Di Maria, C. M. Saluzzo, P. Quaretti, G. Rodolico, P. Scagnelli, et al. A Study on Maximum Skin Dose in Cerebral Embolization Procedures AJNR Am. J. Neuroradiol., March 1, 2007; 28(3): 503 - 507. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Zanzonico, L. N. Rothenberg, and H. W. Strauss Radiation Exposure of Computed Tomography and Direct Intracoronary Angiography: Risk Has its Reward J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1846 - 1849. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Peterzol, E. Quai, R. Padovani, G. Bernardi, C. J. Kotre, and A. Dowling Reference levels in PTCA as a function of procedure complexity Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 54 - 58. [Abstract] [Full Text] [PDF] |
||||
![]() |
J G Albert, F Martiny, A Krummenerl, K Stock, J Lesske, C M Gobel, E Lotterer, H H Nietsch, C Behrmann, and W E Fleig Diagnosis of small bowel Crohn's disease: a prospective comparison of capsule endoscopy with magnetic resonance imaging and fluoroscopic enteroclysis Gut, December 1, 2005; 54(12): 1721 - 1727. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Lickfett, M. Mahesh, C. Vasamreddy, D. Bradley, V. Jayam, Z. Eldadah, T. Dickfeld, D. Kearney, D. Dalal, B. Luderitz, et al. Radiation Exposure During Catheter Ablation of Atrial Fibrillation Circulation, November 9, 2004; 110(19): 3003 - 3010. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Seibert Vertebroplasty and Kyphoplasty: Do Fluoroscopy Operators Know about Radiation Dose, and Should They Want to Know? Radiology, September 1, 2004; 232(3): 633 - 634. [Full Text] [PDF] |
||||
![]() |
K. Perisinakis, J. Damilakis, N. Theocharopoulos, G. Papadokostakis, A. Hadjipavlou, and N. Gourtsoyiannis Patient Exposure and Associated Radiation Risks from Fluoroscopically Guided Vertebroplasty or Kyphoplasty Radiology, September 1, 2004; 232(3): 701 - 707. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Perisinakis, J. Damilakis, N. Theocharopoulos, G. Papadokostakis, A. Hadjipavlou, and N. Gourtsoyiannis Patient effective dose and radiogenic risks from fluoroscopically assisted surgical reconstruction of femoral fractures Radiat Prot Dosimetry, January 1, 2004; 108(1): 65 - 72. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Given, I. C. Thacker, M. D. Baker, and P. P. Morris Fluoroscopy Fade for Embolization of Vein of Galen Malformation AJNR Am. J. Neuroradiol., February 1, 2003; 24(2): 267 - 270. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Miller, S. Balter, P. T. Noonan, and J. D. Georgia Minimizing Radiation-induced Skin Injury in Interventional Radiology Procedures Radiology, November 1, 2002; 225(2): 329 - 336. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOGRAPHICS | RADIOLOGY | RSNA JOURNALS ONLINE |