SCIENTIFIC EXHIBIT |
1 From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. Recipient of a Certificate of Merit award for a scientific exhibit at the 1999 RSNA scientific assembly. Received February 25, 2000; revision requested April 24 and received June 19; accepted June 19. Address correspondence to C.H.M. (e-mail: mccollough.cynthia@mayo.edu).
To visually demonstrate the effective temporal resolution of subsecond conventional (slip-ring) and electron-beam computed tomographic (CT) systems, two phantoms containing high-contrast test objects were scanned with a slip-ring CT system (effective exposure time, 0.5 second) and an electron-beam CT system (exposure time, 0.1 second). Images were acquired of each phantom at rest, during translation along the x axis at speeds of 10100 mm/sec, and during rotation about isocenter at speeds of 0.1 and 0.5 revolution per second. Motion artifacts and loss of spatial resolution were judged to be absent, noticeable, or severe. For 0.5-second conventional CT images, motion artifacts and loss of spatial resolution were noticeable at 10 mm/sec and 0.1 revolution per second and were severe at speeds greater than or equal to 20 mm/sec and at 0.5 revolution per second. For 0.1-second electron-beam CT scans, noticeable, but not severe, motion artifacts and loss of spatial resolution occurred at speeds between 40 and 100 mm/sec and at 0.5 revolution per second. Over the range of physiologic speeds examined, the images provide visually compelling evidence of the effect of improving temporal resolution in CT.
Index Terms: Computed tomography (CT), artifact Computed tomography (CT), electron beam Computed tomography (CT), image quality Phantoms Test objects
This article has been cited by other articles:
![]() |
R. Popilock, K. Sandrasagaren, L. Harris, and K. A. Kaser CT Artifact Recognition for the Nuclear Technologist J. Nucl. Med. Technol., June 1, 2008; 36(2): 79 - 81. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Grosjean, B. Sauer, R. M. Guerra, M. Daudon, A. Blum, J. Felblinger, and J. Hubert Characterization of Human Renal Stones with MDCT: Advantage of Dual Energy and Limitations Due to Respiratory Motion Am. J. Roentgenol., March 1, 2008; 190(3): 720 - 728. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Liu, C. Cuevas, A. A. Moss, O. Kolokythas, T. J. Dubinsky, and P. E. Kinahan Gas Bubble Motion Artifact in MDCT Am. J. Roentgenol., February 1, 2008; 190(2): 294 - 299. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. S. Choi, B. W. Choi, K. O. Choe, D. Choi, K.-J. Yoo, M.-I. Kim, and J. Kim Pitfalls, Artifacts, and Remedies in Multi- Detector Row CT Coronary Angiography RadioGraphics, May 1, 2004; 24(3): 787 - 800. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOGRAPHICS | RADIOLOGY | RSNA JOURNALS ONLINE |